Ventrium-3 is a three channel ECG machine which can simultaneously record 12 leads of ECG and can print ECG report on a thermal printer. A built-in highly accurate interpretation algorithm can generate interpretation reports by analyzing recorded ECG waveforms which not only ease out the diagnostic procedure for general practitioners but also provides detailed analysis to cardiologist for further diagnosis and treatment.
General Features
- Portable, Light weight and Ergonomic Design
- Built-in 3-channel thermal printer
- Simultaneous Acquisition & Display of 12 lead ECG
- Automatic, Manual & Rhythm modes
- Auto Interpretation
- Built-in Storage capacity of atleast 1000 ECGs (with patient’s info) for data review and analysis purpose.
- With ECG Sync Software for Data Transfer and Print
Technical Features
- 5” color LCD slanted display for easy and comfort viewing
- Leads: 12- leads: I, II, III, aVL, aVR, aVF& V1 – V6
- Recording Speed :
- Auto Record: 25 mm/s, 50 mm/s
- Manual Record: 5, 10, 25, 50 mm/s
- Rhythm Record: 25 mm/s, 50 mm/s
- Sensitivity: 5, 10, 20 mm/mV (Accuracy: ±5%)
- Recording Paper: Thermal Paper roll {Size: 80 mm (w) x 20 m (L)}
- Supported Printing Modes: 4*3,3*4, 3*4 + 1, 2*6, 2*6 + 1, 3-2, 3-2 + 1, Rhythm 2, Rhythm 3, Rhythm 4, manual mode
- Measurement Parameters: HR, P-R interval, P Duration, QRS Duration, Q Interval, Q-Tc, P Axis, QRS Axis, T Axis, R(V5), S(V1), R(V5) + S(V1)
- Digital Filters: AC (50/60 Hz), EMG (25/35 Hz) & DFT Filters
- Frequency Response: 0.05 Hz – 150 Hz (-3 dB ~ 0.4 dB)
- Time Constant: Time ≥ 3.2 s
- CMRR: ≥ 60 dB, ≥ 100 dB (with all filters enabled)
- Sampling Rate: 1000 Hz
- Patient Leakage Current: ≤ 10 µA
- Classification: Class I. Type CF applied part
- Input way: Floating and defibrillation protection
- Power Supply:
- AC: 100 – 240 V AC, 50/60 Hz;
- DC: 7.4 V, 3700 mAh rechargeable lithium battery
- Battery Backup: Approx. 1.5 Hours (continuous printing)
- Physical Characteristics:
- Machine Size: 315 (L) x 215 (W) x 77 (H) mm
- Machine Net Weight: ~ 1.7 Kg